3  Programming the Analyzer

STATus Subsystem

Questionable Power Enable

This command determines which bits in the Questionable Power Condition Register will set bits in
the Questionable Power Event register, which also sets the Power Summary bit (bit 3) in the
Questionable Register.

The variable <integer> is the sum of the decimal values of the bits you want to enable.

Mode

All

Remote
Command

:STATus:QUEStionable:POWer:ENABle <integer>

:STATus:QUEStionable:POWer:ENABle?

Example

STAT:QUES:POW:ENAB 32

50 MHz Input Pwr too High for Cal will be reported to the Power Summary of the Status Questionable
register.

Preset

32767

Min

0

Max

32767

Status
Bits/OPC
dependencies

Sequential command

Initial S/W
Revision

Prior to A.02.00

Questionable Power Event Query

This query returns the decimal value of the sum of the bits in the Questionable Power Event
register.

The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.

Mode

All

Remote Command

:STATus:QUEStionable:POWer[:EVENt]?

Example

STAT:QUES:POW?

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00