6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output

Key Path

Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe

Remote Command

:CALibration:IQ:PROBe:I|Q:CLEar

Example

Clear the calibration data for the I channel and the current probe (with EEPROM identification) or
probe type (without EEPROM identification).

:CAL:IQ:PROBe:I:CLE

Initial S/W Revision

Prior to A.02.00

I/Q Cable Calibration Time (Remote Command Only)

Returns the last date and time that the I/Q Cable Calibration was performed for a specific port.
This is a remote query command only.

Remote Command

:CALibration:IQ:FLATness:I|IBAR|Q|QBAR:TIME?

Example

:CAL:IQ:FLAT:I:TIME?

Notes

This returns 6 integer values: year, month, day, hour, minute, second. When no calibration has
been performed, all values will be 0.

Initial S/W Revision

A.02.00

I/Q Probe Calibration Time (Remote Command Only)

Return the last date and time that the I/Q Probe Calibration was performed for a specific port.
This is a remote query command only.

Remote Command

:CALibration:IQ:PROBe:I|IBAR|Q|QBAR:TIME?

Example

:CAL:IQ:PROB:I:TIME?

Notes

This returns 6 integer values: year, month, day, hour, minute, second. When no calibration has
been performed, all values are 0. The value is specific to both the port and probe, so the value will
change as probes are connected or disconnected.

Initial S/W Revision

A.02.00

RF Calibrator

Lets you choose a calibrator signal to look at or turns the calibrator "off".

Key Path

Input/Output

616