Agilent Technologies Signal Analyzer N9030a User Manual

Page 848

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup

De-emphasis (FM Demod only)

The De-emphasis setting controls a single-pole filter (6 dB/octave roll off), usually to counter
intentional pre-emphasis in the transmitter. When De-emphasis state is OFF the hardware digital
filter is bypassed, otherwise the setting is applied

The De-emphasis softkey is only available when FM is the demod selected. It is grayed out for AM
and PM.

Key Path

Meas Setup, Analog Demod Tune & Listen, FM

Remote Command

[:SENSe]:DEMod:FM:DEEMphasis OFF|US25|US50|US75|US750

[:SENSe]:DEMod:FM:DEEMphasis?

Example

DEM:FM:DEEM US75

DEM:FM:DEEM?

Dependencies

Only available in FM. Grayed out for AM and PM.

Preset

US75 (recommended for US commercial FM 75 µs pre-emphasis)

State Saved

Saved in instrument state.

Readback line

1-of-N selection

Initial S/W Revision

Prior to A.02.00

De-emphasis (FM Demod only)

The De-emphasis setting controls a single-pole filter (6 dB/octave roll off), usually to counter
intentional pre-emphasis in the transmitter. When De-emphasis state is OFF the hardware digital
filter is bypassed, otherwise the setting is applied

The De-emphasis softkey is only available when FM is the demod selected. It is grayed out for AM
and PM.

Key Path

Meas Setup, Analog Demod Tune & Listen, FM

Remote Command

[:SENSe]:DEMod:FM:DEEMphasis OFF|US25|US50|US75|US750

[:SENSe]:DEMod:FM:DEEMphasis?

Example

DEM:FM:DEEM US75

DEM:FM:DEEM?

Dependencies

Only available in FM. Grayed out for AM and PM.

Preset

US75 (recommended for US commercial FM 75 µs pre-emphasis)

State Saved

Saved in instrument state.

Readback line

1-of-N selection

Initial S/W Revision

Prior to A.02.00

848

Remote Language Compatibility Measurement Application Reference

Advertising