6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control

Remote Command

:TRIGger[:SEQuence]:RFBurst:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:RFBurst:SLOPe?

Example

TRIG:RFB:SLOP NEG

Couplings

This same slope is used in the RF Burst selection for the trigger source in the Trigger menu and
for the period timer sync source (in the Trigger menu and in the Gate Source menu).

Preset

POSitive

State Saved

Saved in instrument state

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:FRAMe:RFBurst:SLOPe

Backwards Compatibility
Notes

The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.

Initial S/W Revision

Prior to A.02.00

Periodic Timer (Frame Trigger)

Pressing this key, when it is not selected, selects the internal periodic timer signal as the trigger.
Triggering occurrences are set by the Period parameter, which is modified by the Sync Source
and Offset. Pressing this key, when it is already selected, accesses the periodic timer trigger
setup functions.

If you do not have a sync source selected (it is Off), then the internal timer will not be
synchronized with any external timing events.

Key Path

Trigger

Example

TRIG:SOUR FRAM                 Swept SA measurement

TRIG:<meas>:SOUR FRAM   Measurements other than Swept SA

State Saved

Saved in instrument state

Readback

[Sync: <value of Sync Source>], for example, [Sync: External 1]

Status Bits/OPC
dependencies

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping
or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger
event occurs and all the applicable trigger criteria have been met). A corresponding pop-up
message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2
sec. This message goes away when a trigger signal appears.

Initial S/W Revision

Prior to A.02.00

Periodic Timer Triggering:

1214

Remote Language Compatibility Measurement Application Reference

Advertising