Standard event status register query, Identification query – Agilent Technologies Signal Analyzer N9030a User Manual

Page 105

Advertising
background image

3  Programming the Analyzer

IEEE 488.2 Common Commands

Standard Event Status Register Query

Queries and clears the standard event status event register. (This is a destructive read.) The value
returned is a hexadecimal number that reflects the current state (0/1) of all the bits in the
register.

Remote Command

*ESR?

Example

*ESR? Returns a 1 if there is either a query or command error, otherwise it returns a zero.

Notes

For related commands, see the STATus subsystem commands.

Preset

0

Min

0

Max

255

Status Bits/OPC
dependencies

Standard Event Status Register (bits 0 – 7).

Initial S/W Revision

Prior to A.02.00

Identification Query

Returns a string of instrument identification information. The string will contain the model
number, serial number, and firmware revision.

The response is organized into four fields separated by commas. The field definitions are as
follows:

Manufacturer

Model

Serial number

Firmware version

Key Path

No equivalent key. See related key System, Show System.

Remote Command

*IDN?

Example

*IDN? Returns instrument identification information, such as:

Agilent Technologies, N9020A, US01020004, A.01.02

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference

105