Trigger source presets – Agilent Technologies Signal Analyzer N9030a User Manual

Page 1393

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trigger

Other trigger-related commands are found in the INITiate and ABORt SCPI command
subsystems.

*OPC should be used after requesting data. This will hold off any subsequent changes to the
selected trigger source, until after the sweep is completed and the data is returned.

Available ranges and presets can vary from mode to mode.

Dependencies

In some models, there is no second External input.  In these models, the External 2 key is
blanked and the EXTernal2 parameter will generate a “Hardware missing; Not available for this
model number” message.

Preset

See table below

Status Bits/OPC
dependencies

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping
or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger
event occurs and all the applicable trigger criteria have been met). A corresponding pop-up
message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2
sec. This message goes away when a trigger signal appears.

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:SOURCe EXTernal

For backward compatibility, the parameter EXTernal is mapped to EXTernal1

Backwards
Compatibility SCPI

[:SENSe]:<measurement>:TRIGger:SOURce

This backwards compatibility alias command is provided for ESA/PSA compatibility

This backwards compatibility command does not apply to the Swept SA measurement, for that
just use :TRIGger:SOURCe

This backwards compatibility command does not apply to the monitor spectrum, log plot and spot
frequency measurements

Backwards
Compatibility SCPI

[:SENSe]:<measurement>:TRIGger:SOURce IF

In earlier instruments, the parameter IF was used by apps for the video trigger, so using the IF
parameter selects VIDeo triggering. Sending IF in the command causes VID to be returned to a
query.

Backwards
Compatibility SCPI

[:SENSe]:ACPR:TRIGger:SOURce

This backwards Compatibility SCPI command is provided to support the same functionality as
[:SENSe]:ACPr:TRIGger:SOURce (PSA W-CDMA, PSA cdma2000 and PSA 1xEVDO) due to the
fact that the ACPr node conflicts with the ACPower node.

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.03.00

Trigger Source Presets

Here are the Trigger Source Presets for the various measurements:

Remote Language Compatibility Measurement Application Reference

1393