6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Input/Output

Remote Command

:INPut[1]:IQ:BALanced[:STATe] OFF|ON|0|1

:INPut[1]:IQ:BALanced[:STATe]?

Notes

For R&S FSQ-B71 compatibility, with no independent settings for the I and Q channels. Therefore,
it is tied only to the I channel and does not provide an equivalent for the Q channel. For proper
operation of the backwards compatibility command Q Same as I should be set to On.

Preset

OFF

Initial S/W Revision

Prior to A.02.00

I Input Z

Selects the input impedance for the I channel. The impedance applies to both the I and I-bar
ports.

The input impedance controls the hardware signal path impedance match. It is not used for
converting voltage to power. The voltage to power conversion always uses the Reference Z
parameter. The Reference Z parameter applies to both I and Q channels.

Key Path

Input/Output, I/Q, I Setup

Remote Command

:INPut[1]:IQ[:I]:IMPedance LOW|HIGH

:INPut[1]:IQ[:I]:IMPedance?

Example

Set the I channel input impedance to 1 MΩ

INP:IQ:IMP HIGH

Notes

LOW = 50 Ω, HIGH = 1 MΩ

When IQ Path is I+jQ, the I Input Z setting must be the same as the Q Input Z setting.  If the
settings of the two inputs do not match, an error condition message is generated, 159;Settings
Alert;I/Q mismatch:Input Z.

Couplings

Input impedance is a built-in characteristic of a probe. Therefore, whenever a probe is sensed,
this key is disabled and the value is set to match the probe.

When no probe is sensed on Q and Q Same as I is On, the value set for I will also be copied to Q.

Preset

LOW

State Saved

Yes

This is unaffected by a Preset but is set to the default value on a "Restore Input/Output
Defaults" or "Restore System Defaults->All"

Range

50 Ω | 1 MΩ

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference