6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output

Coupling

Some probe types allow coupling to reject low frequencies. This will filter out the DC component of
a signal that is composed of a DC bias plus some AC signal. This control is visible only for probe
types that have this capability.

Key Path

Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe

Remote Command

:INPut:COUPling:I|Q DC|LFR1|LFR2

:INPut:COUPling:I|Q?

Example

Set the probe to low frequency rejection below 1.7 Hz.

INP:COUP:I LFR1

Notes

Only some probe types support Coupling. For those that do, each probe type has its own Coupling
setting. As probes are changed the Coupling value will reflect the new probe's setting. Changing
the Coupling affects only the current probe type's setting and leaves all others unchanged.

Preset

DC

State Saved

Saved with probe calibration data. It survives a power cycle and is not affected by a Preset or
Restore.

Range

DC | AC 1.7 Hz LFR1 | AC 0.14 Hz LFR2

Readback Text

DC | LFR1 | LFR2

Initial S/W Revision

Prior to A.02.00

DC

Turns off low frequency rejection, allowing signals down to DC.

Key Path

Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe, Coupling

Example

Turn off low frequency rejection on the I channel

INP:COUP:I DC

Initial S/W Revision

Prior to A.02.00

LFR1

Turns on low frequency rejection, rejecting signal component lower than 1.7 Hz.

Key Path

Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe, Coupling

Example

Turn on low frequency rejection on the I channel for frequencies lower than 1.7 Hz

INP:COUP:I LFR1

Initial S/W Revision

Prior to A.02.00

614

Remote Language Compatibility Measurement Application Reference

Advertising