6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup

Key Path

Meas Setup

Remote Command

[:SENSe]:AVERage:TYPE:AUTO OFF|ON|0|1

[:SENSe]:AVERage:TYPE:AUTO?

Preset

ON

State Saved

Saved in Instrument State

Readback line

1-of-N selection as

Log-Pwr (Video)

for Log-Pwr (Video) Avg

Pwr (RMS)

for Power Avg

Voltage

for Voltage

Initial S/W Revision

Prior to A.02.00

Remote Command

[:SENSe]:AVERage:TYPE RMS|LOG|SCALar [:SENSe]:AVERage:TYPE?

Notes

Parameters map to avg types as:

RMS = Pwr (RMS) Avg

LOG = Log-Pwr (Video) Avg

SCALar = Voltage Avg

Preset

LOG

Backwards Compatibility
Notes

The following legacy parameters to the [:SENSe]:AVERage:TYPE command are aliased as shown:

LINear aliased to SCALar, sets Scalar averaging

VOLTage aliased to SCALar, sets Scalar averaging

VIDeo aliased to LOG, sets Log-Power averaging

LPOWer aliased to LOG, sets Log-Power averaging

POWer aliased to RMS , sets RMS averaging

Initial S/W Revision

Prior to A.02.00

More Information

When you select log-power averaging, the measurement results are the average of the signal level
in logarithmic units (decibels). When you select power average (RMS), all measured results are
converted into power units before averaging and filtering operations, and converted back to
decibels for displaying. Remember: there can be significant differences between the average of the
log of power and the log of the average power.

These are the averaging processes within a spectrum analyzer and all of them are affected by this
setting:

790

Remote Language Compatibility Measurement Application Reference