6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker Function

If the Measure at Marker window is being displayed, and one of the detectors is changed, any
value being displayed for that detector changes to “---“ until the next successful reading from that
detector.

Key Path

Marker Function, Measure at Marker,

Remote Command

:CALCulate:MAMarker:DETector[1] | 2 | 3 OFF | NORMal | AVERage |
POSitive | SAMPle | NEGative | QPEak | EAVerage | RAVerage

:CALCulate:MAMarker:DETector[1]|2|3?

Example

:

CALC:MAM:DET2 QPE

Sets the detector for measure at marker detector 2 to Quasi peak

:CALC:MAM:DET OFF

Sets the detector for measure at marker detector 1 to Off

State Saved

Saved in instrument state

Initial S/W Revision

A.02.00

Key Path

Marker Function, Measure at Marker,

Remote Command

:CALCulate:MAMarker:DETector[1]|2|3:DWELl <dwell time>

:CALCulate:MAMarker:DETector[1]|2|3:DWELl?

Example

:CALC:MAM:DET2:DWEL 500 ms

Sets the detector for measure at marker detector 2 to dwell for 500 ms

State Saved

Saved in instrument state

Backwards
Compatibility SCPI

[:SENSe]:EMI:MEASure:DETector:DWELl <dwell time>

This command is included for compatibility with the E7400 and PSA option 239 .  Sets all of the
detectors’ dwell times to the specified amount

Initial S/W Revision

A.02.00

Detector 1

This menu lets you select the detector to be used for Detector 1, or turn Detector 1 off.  This is a
1-of-N menu that shows the normal list of detectors, but with the “Auto” key replaced by “Off”.

Key Path

Marker Function, Measure at Marker, Detectors

Remote Command

See

"Detectors" on page 755

. 

Example

:CALC:MAM:DET QPE

Sets the detector for measure at marker detector 1 to Quasi peak

:CALC:MAM:DET OFF

756

Remote Language Compatibility Measurement Application Reference