5  Legacy Command Descriptions

IP (Instrument Preset)

IP (Instrument Preset)

Syntax

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

Performs an instrument preset, setting the instrument back to its factory settings. IP does not
affect the contents of any data or trace registers or stored preselector data. IP does not clear the
input or output data buffers on the 8560-series analyzers, but does clear them on the 8566A/B,
8568A/B.

Instrument preset automatically occurs when you turn on the instrument. IP is a good starting
point for many measurement processes. When IP is executed remotely, the instrument does not
necessarily execute a complete sweep, however. You should execute a

"TS (Take Sweep)" on page

497

to ensure that the trace data is valid after an IP.

N9061A executes this command after any language switch on the X-Series instrument.

Format

IP

Query Data Type

N/A

SCPI Equivalent Commands

None

Notes

The functions of IP are identical to the command

"KST (Fast Preset)" on page 347

.

If the external amplifier gain has been set, executing IP does not reset this value. This is
to protect the instrument.

Remote Language Compatibility Measurement Application Reference