6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker Function

(buckets) by translating back based on the current X Axis Scale settings at the time the query is
sent.

See

"Band Function Backwards Compatibility" on page 731

for more information

Initial S/W Revision

Prior to A.02.00

Band/Interval Right

Sets the right edge frequency or time for the band of the selected marker. The left edge is
unaffected

In the table below, sweep_width = max(1,sweep_points–1) and sweep_points is the number of
sweep points, set in the Sweep menu.

It is legal to change the width of the band even if there is no marker function on. Generally this
can only happen by sending the SCPI command since access to the menu is restricted if no
marker function is on.

Key Path

Marker Function, Band Adjust

Remote Command

:CALCulate:MARKer[1]|2|...12:FUNCtion:BAND:RIGHt <freq>

:CALCulate:MARKer[1]|2|...12:FUNCtion:BAND:RIGHt?

Example

:CALC:MARK12:FUNC:BAND:RIGHt 20 GHz sets the right edge of the band span of marker 12 to
20 GHz

:CALC:MARK:FUNC:BAND:RIGHt? queries the band span of Marker 1

Notes

Units are those of the trace’s domain, Hz for frequency domain, s for time domain. When the
right edge is moved, the left edge stays anchored; thus, the marker’s frequency will change.

Notes

Sending this command selects the subopcoded marker

The unit of the parameter must match the current domain of the trace the selected marker is on,
or an invalid suffix error will be generated. If no unit is sent the fundamental unit for the trace
domain will be used (Hz for freq domain traces, s for time domain traces).

Note that all the values provided in this table are only valid for frequency domain traces. If the
current domain of the trace is time domain, values and unit will be different. In frequency
domain, the Preset value is dependent on the frequency range of the instrument. The default
value 1.3245 GHz is appropriate only if the instrument is a 26.5 GHz instrument (Option 526). In a
26.5 GHz Instrument, the default span is 26.49 GHz, so 5% of the span corresponds to 1.3245
GHz.

Couplings

Changing the Band/Interval Right necessarily changes the Band/Interval Span and
Band/Interval Center values

Band/Interval Span is set to 5% of span when any marker function is turned on if and only if it is
zero at that time

748

Remote Language Compatibility Measurement Application Reference

Advertising