6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Recall

recalled.

Once the state is loaded, the trace data must be loaded. The internal flags are consulted to see
which trace to load and the "To Trace" setting to see where to load it. Trace data is always
loaded with the specified trace set to View, so that the data is visible and not updating(so as not
to erase the recalled data). If the file is an "all trace" file, all traces are loaded with the saved data
(to the original trace the data was saved from) and set to View. Traces whose data is not loaded
are restored to the update state that existed when they were saved.

After the Recall the analyzer exits the Recall menu and returns to the previous menu.

Some modes and measurements do not have available all 6 traces. Phase Noise mode command,
for example, is: MMEMory:LOAD:TRACe TRACE1|TRACE2|TRACE3,<filename>

Initial S/W Revision

Prior to A.02.00

To Trace

These menu selections let you choose the Trace where the recalled saved trace will go. Not all
modes have the full 6 traces available.  The default is the currently selected trace, selected in this
menu or in the Trace/Detector, Export Data, Import Data, or Save Trace menus, except if you have
chosen All, then it remains chosen until you specifically change it to a single trace.

If the .trace file is an "all trace" file, "To Trace" is ignored and the traces each go back to the trace
from which they were saved.

Once selected, the key returns back to the Recall Trace menu and the selected Trace number is
annotated on the key. Now you have selected exactly where the trace needs to be recalled.To
trigger a recall of the selected Trace, you must select the Open key in the Recall Trace menu.

When you select a trace, it makes that trace the current trace, so it displays on top of all of the
other traces.

Key Path

Save, Data, Trace

Mode

SA

Initial S/W Revision

Prior to A.02.00

To Trace

These menu selections let you choose the Trace where the recalled saved trace will go. Not all
modes have the full 6 traces available.  The default is the currently selected trace, selected in this
menu or in the Trace/Detector, Export Data, Import Data, or Save Trace menus, except if you have
chosen All, then it remains chosen until you specifically change it to a single trace.