6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger

Key Path

Trigger, Auto/Holdoff

Remote Command

:TRIGger[:SEQuence]:HOLDoff <time>

:TRIGger[:SEQuence]:HOLDoff?

:TRIGger[:SEQuence]:HOLDoff:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:HOLDoff:STATe?

Example

TRIG:HOLD:STAT ON

TRIG:HOLD 100 ms

Dependencies

Unavailable if the selected Input is BBIQ.  If this is the case, the key is grayed out if it is pressed
the informational message “Feature not supported for this Input” is displayed.  If the SCPI
command is sent, the error “Settings conflict; Feature not supported for this Input” is generated.

Preset

Off, 100 ms

State Saved

Saved in instrument state

Min

0 s

Max

0.5 s

Default Unit

s

Initial S/W Revision

Prior to A.02.00