6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Source

Source Frequency = (Analyzer Frequency *Multiplier Numerator / Multiplier Denominator) +
Source Frequency Offset

Analyzer Frequency is the frequency to which the analyzer is set, which is the analyzer’s
displayed frequency, offset by any Freq Offset set under the Frequency hardkey.  Source Frequency
Offset is the value set under Source, Frequency, Freq Offset.

Key Path

Source

Readback

none in Tracking Source mode

Initial S/W Revision

A.06.01

Multiplier Numerator

The multiplier numerator parameter offsets the source frequency from the analyzer frequency. 
The source frequency tracks the SA frequency according to the source frequency equation shown
under the Source, Frequency key description.

The multiplier numerator must be restricted to operate within the range of the source minimum
and maximum frequencies.

Key Path

Source, Frequency

Mode

SA

Scope

Mode Global

Remote Command

:SOURce:FREQuency[:MULTiplier]:NUMerator <integer>

:SOURce:FREQuency[:MULTiplier]:NUMerator?

Example

:SOUR:FREQ:NUM 3

Sets the source frequency multiplier numerator to 3.

Dependencies

If the currently selected source does not support this capability (for example, an internal
Tracking Generator which must track the LO), this key is forced to its Preset value and grayed
out

Preset

This is unaffected by Mode Preset but is set to 1 on a Source Preset or Restore Input/Output
Defaults.

State Saved

Part of the Input/Output system, which means it is Loaded and Saved with state.

Min

1

Max

1000

Initial S/W Revision

A.06.01