6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Input/Output

Example

:CORR:CSET:ANT GAUS

Readback

" dBG"

Initial S/W Revision

A.02.00

dBµA

Sets the antenna unit to dBµA. If this correction is turned on, and Apply Corrections is on, the Y
Axis Unit will then be forced to dBµA and all other Y Axis Unit selections will be grayed out.

Key Path

Input/Output, Corrections, Properties, Antenna Unit

Example

:CORR:CSET:ANT UA

Readback

" dBµA"

Initial S/W Revision

A.11.00

Frequency Interpolation

This setting controls how the correction values per-bucket are calculated. We interpolate between
frequencies in either the logarithmic or linear scale.

This setting is handled and stored individually per correction set.

See

"Interpolation" on page 633

Key Path

Input/Output, Corrections, Properties

Remote Command

[:SENSe]:CORRection:CSET[1]|2|...6:X:SPACing LINear | LOGarithmic

[:SENSe]:CORRection:CSET[1]|2|...6:X:SPACing?

Example

CORR:CSET:X:SPAC LIN

Preset

Unaffected by a Preset. Set to Linear by Restore Input/Output Defaults.

State Saved

Saved in instrument state.

Initial S/W Revision

A.02.00

Interpolation

For each bucket processed by the application, all of the correction factors at the frequency of
interest (center frequency of each bucket) are summed and added to the amplitude. All trace
operations and post processing treat this post-summation value as the true signal to use.