6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Mode

following:

:INST:SEL EMI
:CONF FSC

This results in the instrument being placed in the EMI Receiver Mode, running the Frequency
Scan measurement, in order to emulate the ESU Receiver Mode.

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.10.01

Example

:INST ‘SA’

Notes

The query is not a quoted string. It is an enumeration as indicated in the Instrument Select table
above.

The command must be sequential: that is, continued parsing of commands cannot proceed until
the instrument select is complete and the resultant SCPI trees are available.

Backwards
Compatibility SCPI

:INSTrument[:SELect] ‘SA’|’PNOISE’|’EDGE’|’GSM’|’BASIC’

Initial S/W Revision

Prior to A.02.00

More Information

The Mode name appears on the banner after the word “Agilent” followed by the Measurement
Title.  For example, for the Spectrum Analyzer mode with the Swept SA measurement running
displays:

It is possible to specify the order in which the Modes appear in the Mode menu, using the
Configure Applications utility (System, Power On, Configure Applications).  It is also possible, using
the same utility, to specify a subset of the available applications to load into memory at startup
time, which can significantly decrease the startup time of the analyzer. During runtime, if an
application that is not loaded into memory is selected (by either pressing that application's Mode
key or sending that application's :INST:SEL command over SCPI), there will be a pause while the
application is loaded.  During this pause a message box appears stating “Loading application,
please wait…”.

Each application (Mode) that runs in an X-Series instrument consumes virtual memory.  The
various applications consume varying amounts of virtual memory, and as more applications run,
the memory consumption increases. Once an application has been loaded, some of its memory

Remote Language Compatibility Measurement Application Reference