Agilent Technologies Signal Analyzer N9030a User Manual

Page 142

Advertising
background image

3  Programming the Analyzer
STATus Subsystem

dependencies

Initial S/W Revision

Prior to A.02.00

Operation Event Query

This query returns the decimal value of the sum of the bits in the Operation Event register.

The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.

Mode

All

Remote Command

:STATus:OPERation[:EVENt]?

Example

STAT:OPER?

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Operation Negative Transition

This command determines which bits in the Operation Condition register will set the
corresponding bit in the Operation Event register when the condition register bit has a negative
transition (1 to 0). The variable <integer> is the sum of the decimal values of the bits that you
want to enable.

Mode

All

Remote Command

:STATus:OPERation:NTRansition <integer>

:STATus:OPERation:NTRansition?

Example

STAT:OPER:NTR 1 Align Now operation complete will be reported to the Status Byte Register.

Preset

0

Min

0

Max

32767

Status Bits/OPC
dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

142

Remote Language Compatibility Measurement Application Reference

Advertising