6  RLC Swept SA Measurement Front-Panel & SCPI Reference
System

No confirmation is required when Alert is configured through a remote command.

Key Path

System, Alignments, Auto Align, Alert

Mode

All

Example

:CAL:AUTO:ALER NONE

Initial S/W Revision

Prior to A.02.00

Execute Expired Alignments (Remote Command Only)

Alignments can be expired in the situation where Auto Align is in the state of Partial or Off. This
feature runs the alignments that have expired. This is different than performing an Align All, Now
operation, which performs an alignment of all subsystems, regardless of whether they are needed
or not. Execute Expired Alignments performs alignments of only the individual subsystems that
have become due.

Mode

All

Remote Command

:CALibration:EXPired?

Example

:CAL:EXP?

Notes

:CALibration:EXPired? returns 0 if successful

:CALibration:EXPired? returns 1 if failed

Initial S/W Revision

Prior to A.02.00

Align Now

Accesses alignment processes that are immediate action operations. They perform complete
operations and run until they are complete.

1268

Remote Language Compatibility Measurement Application Reference