6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Sweep/Control

l

When Gate is on, Offset Res BW and Offset Video BW are ignored (if you set these values)

and the measurement works as if all Offset Res BW and all Offset Video BW are coupled

with the Res BW and the Video BW under the BW menu. When Gate is on, the Offset BW

key in the Offset/Limit menu is grayed out.

Preset

Off

State Saved

Saved in instrument state

Range

On|Off

Backwards
Compatibility SCPI

[:SENSe]:SWEep:TIME:GATE[:STATe]

ESA compatibility

Backwards Compatibility
Notes

In ESA, Trig Delay (On) and Gate (On) could not be active at the same time. This dependency
does not exist in PSA or in X-Series.

Initial S/W Revision

Prior to A.02.00

Gate View On/Off

Turning on Gate View in the Swept SA measurement provides a single-window gate view display..

Turning on Gate View in other measurements shows the split-screen Gate View. In these
measurements, when the Gate View is on, the regular view of the current measurement traces and
results are reduced vertically to about 70% of the regular height. The Zero Span window, showing
the positions of the Gate, is shown between the Measurement Bar and the reduced measurement
window. By reducing the height of the measurement window, some of the annotation on the Data
Display may not fit and is not shown.

Key Path

Sweep/Control, Gate

Remote Command

[:SENSe]:SWEep:EGATe:VIEW ON|OFF|1|0

[:SENSe]:SWEep:EGATe:VIEW?

Example

SWE:EGAT:VIEW ON turns on the gate view.

Dependencies

In the Swept SA measurement:

In Gate View, the regular Sweep Time key is grayed out . When pressed, the grayed out key puts
up the informational message "Use Gate View Sweep Time in the Gate menu."

In the other measurements:

When you turn Gate View on, the lower window takes on the current state of the instrument.
Upon leaving Gate View, the instrument takes on the state of the lower window.

When you turn Gate View on, the upper window Sweep Time is set to the gate view sweep time.

Couplings

These couplings apply to the Swept SA measurement:

Remote Language Compatibility Measurement Application Reference

1197