3  Programming the Analyzer

STATus Subsystem

Remote Command

:STATus:QUEStionable:CALibration:EXTended:NEEDed:CONDition?

Example

STAT:QUES:CAL:EXT:NEED:COND?

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Calibration Extended Needed Enable

This command determines which bits in the Questionable Calibration Extended Needed Condition
Register will set bits in the Questionable Calibration Extended Needed Event register, which also
sets bit 14 of the Questionable Calibration Register. The variable <integer> is the sum of the
decimal values of the bits you want to enable.

Mode

All

Remote Command

:STATus:QUEStionable:CALibration:EXTended:NEEDed:ENABle <integer>

:STATus:QUEStionable:CALibration:EXTended:NEEDed:ENABle?

Example

STAT:QUES:CAL:EXT:NEED:ENAB 2

Can be used to query if an EMI conducted alignment is needed.

Preset

32767

Min

0

Max

32767

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Calibration Extended Needed Event Query

This query returns the decimal value of the sum of the bits in the Questionable Calibration
Extended Needed Event register.

The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.

Mode

All

Remote Command

:STATus:QUEStionable:CALibration:EXTended:NEEDed[:EVENt]?

Example

STAT:QUES:CAL:EXT:NEED?

Remote Language Compatibility Measurement Application Reference