6  RLC Swept SA Measurement Front-Panel & SCPI Reference
AMPTD Y Scale

COMBined

[:SENSe]:POWer[:RF]:RANGe:OPTimize:ATTenuation?

Notes

The SCPI parameter ELECtrical sets this function to On in single attenuator models.  The SCPI
parameter COMBined is mapped to ELECtrical in single attenuator models; if you send
COMBined, it sets the function to On and returns ELEC to a query.

Dependencies

This key only appears in Dual Attenuator models with an Electronic Attenuator installed. It does
not appear in models with the Single Attenuator configuration, as in the single attenuator
configuration there is no “electronic attenuator” there is only a single integrated attenuator
(which has both a mechanical and electronic stage). 

When Enable Elec Atten is off or grayed out, the Pre-Adjust for Min Clip key is grayed out.

Preset

OFF for Swept SA measurement; ON for all other measurements that support Pre-Adjust for Min
Clip

State Saved

Saved in instrument state

Range

Dual attenuator models:

Off | Elec Atten Only | Mech + Elec Atten

Single attenuator models:

Off | On

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.03.00

Remote Command

[:SENSe]:POWer[:RF]:RANGe:AUTO ON|OFF|1|0

[:SENSe]:POWer[:RF]:RANGe:AUTO?

Notes

ON aliases to "Elec Atten Only" (:POW:RANG:OPT:ATT ELEC)

OFF aliases to "Off" (:POW:RANG:OPT:ATT OFF)

The query :POW:RANG:AUTO? returns true if :POW:RANG:OPT:ATT is not "Off"

Initial S/W Revision

Prior to A.02.00

Off

Turns Pre-Adjust for Min Clip off. This is the default setting.

This key is grayed out in measurements that do not support this functionality. The spectrum
analyzer measurement, Swept SA, does not support this functionality.

Key Path

AMPTD Y Scale, Attenuation, Pre-Adjust for Min Clip

526

Remote Language Compatibility Measurement Application Reference