6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup

Set the Detector to AVERAGE

Set the sweep time to 16ms - sets the variance correctly for good results.

Set a Band/Interval Power Marker function and set the interval over the full width of trace i.e.

Left to 0s and Right to 16ms

Noise Source

This menu allows you to turn the noise source power on or off when making manual noise figure
measurements. 

See

"More Information" on page 856

.

Key Path

Meas Setup

Remote Command

:SOURce:NOISe:TYPE NORMal | SNS

:SOURce:NOISe:TYPE?

Example

SOUR:NOIS:TYPE NORM

Couplings

If no SNS is connected, this parameter will be set to “Normal”

When Type is set to “SNS” and the SNS is disconnected, this parameter gets bumped to
“Normal”

When an SNS is not connected, the SNS type will be grayed (disabled).

Preset

Normal

State Saved

Saved in instrument state.

Range

Normal | SNS

Backwards Compatibility
Notes

In previous Noise Figure analysis applications, this command

could optionally be preceded with the :SENSe keyword.  The

optional :SENSe keyword is no longer supported.

Initial S/W Revision

Prior to A.02.00

More Information

There are 2 types of noise sources: a Smart Noise Source (SNS), and a "Normal" noise source - e.g.
346 series. This menu allows the user to control both. The SNS has its own connector on the rear
of the analyzer and when it is connected the user can then select it from the “Type” 1 of N,
allowing the State parameter to then control the SNS. The "Normal" source is controlled by a BNC
connector that supplies 28V. If SNS is NOT connected then the “state” parameter controls the
"Normal" noise source 28V BNC port. If both are connected the “Type” parameter will determine
which source the “State” parameter will control. Two sources can never be controlled together. The
“SNS attached” SCPI query detailed below can be used remotely to determine if an SNS is
connected. SNS functionality is limited to turning on and off only. The SNS ENR data and
temperature cannot be queried, unless the Noise Figure application is installed. The SNS ENR