6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Sweep/Control

3. If the new absolute RF Burst level differs from the previous by more than 0.5 dB, the new level

is sent to the hardware; otherwise it is not updated (to avoid slowing down the acquisition)

Steps 2 and 3 repeat for subsequent measurements.

Key Path

Trigger, RF Burst

Scope

Meas Global

Remote Command

:TRIGger[:SEQuence]:RFBurst:LEVel:RELative <rel_ampl>

:TRIGger[:SEQuence]:RFBurst:LEVel:RELative?

Example

TRIG:RFB:LEV:REL –10 dB

sets the trigger level of the RF burst envelope signal to the relative level of –10 dB

Notes

Sending this command does not switch the setting from absolute to relative; to switch it you
need to send the :TRIGger[:SEQuence]:RFBurst:LEVel:TYPE command, above.

The relative trigger level is not available in some measurements.  In those measurements the
RELative parameter, and the :TRIGger[:SEQuence]:RFBurst:LEVel:TYPE
command (above), will generate an error if sent.

Dependencies

This key is grayed out and Absolute Trigger Level selected if the required hardware is not present
in your analyzer and the current measurement does not support Relative triggering. 

Preset

–6 dB

GSM:  –25 dB

State Saved

Saved in instrument state

Min

–45 dB

Max

0 dB

Default Unit

dB or dBc

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:RFBurst:LEVel

This legacy command is aliased to :TRIGger[:SEQuence]:RFBurst:LEVel:RELative because the
PSA had ONLY relative burst triggering

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.04.00

Trigger Slope

It is set positive to trigger on a rising edge and negative to trigger on a falling edge.