5  Legacy Command Descriptions

DET (Detection Mode)

DET (Detection Mode)

Syntax

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

Selects the type of instrument detection (NEGative peak, NoRMal, POSitive peak, or SaMPle).

NEG

Enables negative peak detection.

NRM

Enables the ‘rosenfell’ detection algorithm that selectively chooses between positive and negative
values.

POS

Enables positive-peak detection, which displays the maximum video signal detected over a number of
instantaneous samples for a particular frequency.

SMP

Enables sample detection, which uses the instantaneous video signal value. Video averaging and noise-
level markers, when activated, activate sample detection automatically.

Format

DET NEG|NRM|POS|SMP (For option descriptions, see table above)

DET?

Query Data Type

NEG|NRM|POS|SMP

SCPI Equivalent Commands

[:SENSe]:DETector[:FUNCtion] (NEGative|NORMal|POSitive|SAMPle)

[:SENSe]:DETector[:FUNCtion]?

(See

"Detector" on page 1347

)

Preset

NRM

Notes

The functions of DET are identical to the commands

"KSa (Normal Detection)" on page

321

,

"KSb (Positive Peak Detection)" on page 323

,

"KSd (Negative Peak Detection)" on

page 327

, and

"KSe (Sample Detection)" on page 329

.

Remote Language Compatibility Measurement Application Reference

273