6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Sweep/Control

A sample of the Gate View screen in other measurements is shown in the following graphic . This
example is for the ACP measurement:

Turning Gate View off returns the analyzer to the Normal measurement view.

In the Swept SA, the normal measurement view is the single-window Swept SA view. When
returning to this view, the Swept SA measurement returns to the Span it was in before entering
Gate View (even if that is Zero Span).

The Gate View window is triggered from the Gate Source, with zero trigger delay. Also, when
updating the Gate View window, the Gate itself must not operate. So it is internally shut off while
the gate view window is being updated. For the Swept SA measurement, this means that the Gate
is internally shut off whenever the gate view window is displayed. The measurement bar and
softkeys continue to show the Trigger source for the main sweep window and give no indication
that the Gate is shut off or that the Gate View window is triggered from the Gate Source.

When in Gate View, vertical lines are displayed in the Gate View window as follows:

Green lines  are displayed at the gate edges as follows: in Edge Gate, a line is shown for Delay

and one for the end of the Gate period (defined by Length, even in FFT. In Level Gate a line is
shown only for Delay. You can adjust the position of the green lines by adjusting the gate
length and the gate delay. These lines update in the Gate View window as the active function

Remote Language Compatibility Measurement Application Reference