6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Meas Setup

The table editor will only operate properly if the analyzer is sweeping, because its updates are tied to the
sweep system.  Thus, you should not try to use the editor in single sweep, and it will be sluggish during
compute-intensive operations like narrow-span FFT sweeps.

When exiting the edit menu (by using the Return key or by pressing an instrument front panel
key), the editor window turns off, however the Limit is still on and displayed,

and the

amplitude scale remains Log.

Limits are turned off by a Preset, but the Limits arrays (data) are only reset (deleted) by Restore
Mode Defaults.  They survive shutdown and restarting of the analyzer application, which means
they will survive a power cycle.

Key Path

Meas Setup, Limits

Couplings

A remote user can enter or access limit line data via :CALCulate:LLINe[1]|2|3|4|5|6:DATA

Initial S/W Revision

A.02.00

Navigate

Lets you move through the table to edit the desired point

Key Path

Meas Setup, Limits, Edit

Notes

There is no value readback on the key

Min

1

Max

2000

Initial S/W Revision

A.02.00

Frequency

Lets you edit the frequency of the current row.

Key Path

Meas Setup, Limits, Edit

Notes

There is no value readback on the key

Min

0

Max

1 THz

Initial S/W Revision

A.02.00

Remote Language Compatibility Measurement Application Reference