Agilent Technologies Signal Analyzer N9030a User Manual

Page 606

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output

This is unaffected by Preset but is set to the default value on a "Restore Input/Output Defaults"
or "Restore System Defaults->All"

Range

50 Ω | 1 MΩ

Initial S/W Revision

Prior to A.02.00

Q Skew

Sets the skew factor for the Q channel. The skew will shift the channel's data in time. Use this to
compensate for differences in the electrical lengths of the input paths due to cabling and probes.

Key Path

Input/Output, I/Q, Q Setup

Remote Command

[:SENSe]:CORRection:IQ:Q:SKEW <seconds>

[:SENSe]:CORRection:IQ:Q:SKEW?

Example

Delay the data for the Q channel by 10 ns.

CORR:IQ:Q:SKEW 10 ns

Preset

0

State Saved

Yes

This is unaffected by a Preset but is set to the default value on a "Restore Input/Output
Defaults" or "Restore System Defaults->All"

Range

0 s to 100 ns

Min

0 s

Max

+100 ns

Initial S/W Revision

Prior to A.02.00

Q Probe

Accesses the probe setup parameters for the Q channel. See

"I/Q Probe Setup" on page 611

.

Key Path

Input/Output, I/Q, Q Setup

State Saved

No

Readback Text

[<Q port probe id>]

This is reporting the type of probe sensed on the Q port. There is no parameter for overriding
what is sensed.

Initial S/W Revision

Prior to A.02.00

606

Remote Language Compatibility Measurement Application Reference