5  Legacy Command Descriptions

FDSP (Frequency Display Off)

FDSP (Frequency Display Off)

Syntax

Legacy Products

8560 series

Description

Turns the frequency annotation OFF.

Format

FDSP OFF

FDSP?

Query Data Type

‘1’ or ‘0’, indicating ON or OFF.

SCPI Equivalent Commands

See

"ANNOT (Annotation)" on page 234

.

Preset

ON

Couplings

It is not possible enable or disable the frequency annotation alone, leaving other annotation
unaffected. Thus, the FDSP command behaves in the same way as

"ANNOT

(Annotation)" on page 234

. If the FDSP command has been used to disable the frequency

annotation, sending the command ANNOT ON does not re-enable the display annotation.
The display annotation is only enabled by sending the command

"IP (Instrument Preset)"

on page 315

.

Remote Language Compatibility Measurement Application Reference