6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Marker Function

Preset

If 0, Band/Interval Span is set to 5% of span, when a marker function is turned on, which affects
Band/Interval Right

State Saved

Saved in instrument state

Min

0 Hz

Max

Infinity.  Unlike legacy analyzers, where the markers were forced to be on screen, X-Series
marker values are not limited and do not clip

Backwards
Compatibility SCPI

:CALCulate:MARKer[1]|2|...4:X:STOP

See

"Band Function Backwards Compatibility" on page 731

Initial S/W Revision

Prior to A.02.00

Remote Command

:CALCulate:MARKer[1]|2|...4:X:POSition:STOP <integer>

:CALCulate:MARKer[1]|2|...4:X:POSition:STOP?

Preset

1000, the actual value is dependent on the selected number of sweep points.

Backwards
Compatibility SCPI

The legacy command,

:CALCulate:MARKer[n]:X:POSition:STOP <param>

was used to control the Delta marker in trace points (buckets) in Band Pair/Delta Pair mode.
There is no new command for setting the stop of a Band Function in trace points. So, when this
command is received, the analyzer first converts the specified span in trace points to the current
X Axis Scale Units (e.g., frequency or time) of the trace upon which the marker resides. Then,
that value is sent to the

:CALC:MARKer[n]:FUNCtion:BAND:RIGHt <param>

command to set the stop of the marker’s Band Function.

The query form of the command will return the marker function RIGHt value in trace points
(buckets) by translating back based on the current X Axis Scale settings at the time the query is
sent.

See

"Band Function Backwards Compatibility" on page 731

for more information

Initial S/W Revision

Prior to A.02.00

Band Span Auto/Man

Determines whether the Band Span for Marker Noise will track the analyzer’s Span.

Band Span is initialized as specified above, under Band/Interval Span.  Subsequently, if the
analyzer’s Span is changed, the effect on Band Span depends on the Auto/Man setting of Band
Span: