Mkr->cf step, Mkr->start – Agilent Technologies Signal Analyzer N9030a User Manual

Page 779

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Marker To

Mkr->CF Step

Sets the center frequency (CF) step size of the analyzer to the marker frequency, or in a delta-
marker mode, to the frequency difference between the delta and reference markers.

If the currently selected marker is not on when this key is pressed, it will be turned on at the
center of the screen as a normal type marker.

Key Path

Marker ->

Remote Command

:CALCulate:MARKer[1]|2|...12[:SET]:STEP

Example

CALC:MARK1:STEP sets the CF step to the value (or delta value) of marker 1.

Notes

Sending this command selects the subopcoded marker

If specified marker is off, this command will turn it on at the center of the screen as a normal
type marker.

Dependencies

This function is not available (key is grayed out) when x-axis is the time domain

Couplings

All the usual couplings associated with setting CF Step apply.

Initial S/W Revision

Prior to A.02.00

Mkr->Start

Changes the start frequency to the frequency of the selected marker. The marker stays at this
frequency, so it moves to the left edge of the display. In delta marker mode, this function sets the
start frequency to the x-axis value of the delta marker.

If the currently selected marker is not on when this key is pressed, it will be turned on at the
center of the screen as a normal type marker.

Key Path

Marker ->

Remote Command

:CALCulate:MARKer[1]|2|...12[:SET]:STARt

Example

CALC:MARK1:STAR sets the start frequency to the value (or delta value) of marker 1.

Notes

Sending this command selects the subopcoded marker

If specified marker is off, this command will turn it on at the center of the screen as a normal
type marker.

Dependencies

This function is not available (key is grayed out) when x-axis is the time domain

Couplings

All the usual couplings associated with setting Start Frequency apply.

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference

779