6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trigger

External 1

Pressing this key, when it is not selected, selects the external input port that you will use for the
periodic trigger synchronization. Pressing this key, when it is already selected, accesses the
external 1 sync source setup menu.

Key Path

Trigger, Periodic Timer, Sync Source

Example

TRIG:FRAM:SYNC EXT

Couplings

Same as External 1 trigger source.

Readback

External 1

Initial S/W Revision

Prior to A.02.00

Trigger Level

Sets the value where the external 1 trigger input will trigger a new sweep/measurement.

Key Path

Trigger, External 1

Remote Command

:TRIGger[:SEQuence]:EXTernal1:LEVel <level>

:TRIGger[:SEQuence]:EXTernal1:LEVel?

Example

TRIG:EXT1:LEV 0.4 V

Couplings

This same level is used for the Ext1 trigger source in the Trigger menu, for the Ext1 selection in
the Periodic Timer sync source (in the Trigger menu and in the Gate Source menu), and also for
the Ext1 selection in the Gate Source menu.

Preset

1.2 V

State Saved

Saved in instrument state

Min

–5 V

Max

5 V

Default Unit

V

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:EXTernal:LEVel

For backward compatibility, the parameter EXTernal is mapped to EXTernal1

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:FRAMe:EXTernal1:LEVel

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference