6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Peak Search

The Peak Table is updated after each sweep. The list of peaks in the Peak Table can be ordered
either by ascending frequency or by descending amplitude. In either case, the entire trace is first
evaluated and the 20 highest peaks are selected for inclusion in the list. After the peaks are
selected, they are then sorted and displayed according to the Peak Sort setting.

Key Path

Peak Search

Initial S/W Revision

Prior to A.02.00

Peak Table On/Off

Turns Peak Table on/off. When turned on, the display is split into a measurement window and a
peak table display window.

Turning the Peak Table on turns the Marker Table off and vice versa.

Key Path

Peak Search, Peak Table

Remote Command

:CALCulate:MARKer:PEAK:TABLe:STATe OFF|ON|0|1

:CALCulate:MARKer:PEAK:TABLe:STATe?

Example

CALC:MARK:PEAK:TABL:STAT ON Turns on and displays the peak table.

Dependencies

When the Peak Table turns on, if Peak Threshold is On then it becomes the active function.

Preset

OFF

State Saved

Saved in instrument state.

Initial S/W Revision

Prior to A.02.00

Peak Sort

Sets the peak table sorting routine to list the peaks in order of descending amplitude or ascending
frequency. The remote command can also be used to sort the peaks found using the
:CALCulate:DATA:PEAKs command.

Key Path

Peak Search, Peak Table

Remote Command

:CALCulate:MARKer:PEAK:SORT FREQuency|AMPLitude

:CALCulate:MARKer:PEAK:SORT?

Example

CALC:MARK:PEAK:SORT AMPL Sets sorting routine to list peaks in order of descending
amplitude.

CALC:MARK:PEAK:SORT?

Preset

AMPLitude