6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output

Key Path

Input/Output, RF Calibrator

Example

:FEED:AREF REF50

Readback

50 MHz

Initial S/W Revision

Prior to A.02.00

4.8 GHz

Selects the 4.8 GHz internal reference as the input signal.

Key Path

Input/Output, RF Calibrator

Example

:FEED:AREF REF4800

Dependencies

The 4.8 GHz internal reference is only available in some models and frequency range options.  If
the 4.8 GHz reference is not present, the 4.8 GHz softkey will be blanked, and if the REF4800
parameter is sent, the analyzer will generate an error. 

Readback

4.8 GHz

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.03.00

Off

Switches the input back to the selected input (RF or I/Q)

Key Path

Input/Output, RF Calibrator

Example

:FEED:AREF OFF

Readback

Off

Initial S/W Revision

Prior to A.02.00

External Gain

Compensates for gain or loss in the measurement system outside the spectrum analyzer. The
External Gain is subtracted from the amplitude readout (or the loss is added to the amplitude
readout). So, the displayed signal level represents the signal level at the output of the device-
under-test, which can be the input of an external device that provides gain or loss.

618