6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control

This command sets gate parameter values to the ESA preset values, as follows:

l

Gate trigger type = edge

l

Gate polarity = positive

l

Gate delay = 1 us

l

Gate length = 1 us

Remote Command

[:SENSe]:SWEep:TIME:GATE:PRESet 

ESA Compatibility

Initial S/W Revision

Prior to A.02.00

Gate Level (Remote Command Only)

Sets the gate input transition point level for the external TRIGGER inputs on the front and rear
panel. This is a legacy command for PSA compatibility. It is simply an alias to the equivalent
trigger level command.

Remote Command

[:SENSe]:SWEep:EGATe:EXTernal[1]|2:LEVel <voltage>

[:SENSe]:SWEep:EGATe:EXTernal[1]|2:LEVel?

Notes

This command is simply an alias to

:TRIGger[:SEQuence]:EXTernal[1]|2:LEVel

For details refer

Initial S/W Revision

Prior to A.02.00

Gate Polarity (Remote Command Only)

Sets the polarity for the gate signal. This setup is now done using the gate trigger’s slope setting.

l

When Positive (Pos) is selected, a positive-going edge (Edge) or a high voltage (Level) will
satisfy the gate condition, after the delay set with the Gate Delay key.

l

When Negative (Neg) is selected, a negative-going edge (Edge) or a low voltage (Level) will
satisfy the gate condition after the delay.

Remote Command

[:SENSe]:SWEep:EGATe:POLarity NEGative|POSitive

[:SENSe]:SWEep:EGATe:POLarity?

Example

SWE:EGAT:POL NEG

1236

Remote Language Compatibility Measurement Application Reference