Agilent Technologies Signal Analyzer N9030a User Manual

Page 1301

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

System

Factory

This key selects the factory setting, for example:

“Agilent Technologies,N9020A,MY00012345,A.05.01”

where the fields are manufacturer, model number, serial number, firmware revision.

Key Path

System, I/O Config, IDN Response

Example

:SYST:IDN "" null string, restores the factory setting

Initial S/W Revision

A.06.0

User

This key allows you to specify your own response to the *IDN? query.  You may enter your
desired response with the Alpha Editor or a plugin PC keyboard.

When you press this key, the active function becomes the current User string with the cursor at
the end.  This makes it easy to edit the existing string. 

If you enter a null string (for example, by clearing the User String while editing and then pressing
Done

) the analyzer automatically reverts to the Factory setting.

Key Path

System, I/O Config, IDN Response

Example

:SYST:IDN “XYZ Corp, Model 12, 012345, A.01.01”  user specified response

Initial S/W Revision

A.06.00

Query USB Connection (Remote Command Only)

Enables you to determine the speed of the USB connection.

Mode

All

Remote Command

:SYSTem:COMMunicate:USB:CONNection?

Example

:SYST:COMM:USB:CONN?

Notes

l

NONE – Indicates no USB connection has been made.

l

LSPeed – Indicates a USB low speed connection (1.5 Mbps).

Remote Language Compatibility Measurement Application Reference

1301