Agilent Technologies Signal Analyzer N9030a User Manual

Page 576

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
FREQ Channel

Option

when turn TG
on and can’t
tune below
while TG on)

clipped to this
Freq when TG
turned on

above) while TG
on

T03

9 kHz

3.0 GHz

3.08 GHz

T06

9 kHz

6.0 GHz

6.05 GHz

The following table shows the Center Frequency Presets for modes other than Spectrum Analyzer:

Mode

CF Preset for RF

WCDMA

1 GHz

WIMAXOFDMA,

1 GHz

BASIC

1 GHz

ADEMOD

1 GHz

VSA

1 GHz

TDSCDMA

1 GHz

PNOISE

1 GHz

LTE

1 GHz

LTETDD

1 GHz

MSR

1 GHz

GSM

935.2 MHz

NFIGURE

1.505 GHz

RF Center Freq

SCPI command for specifying the RF Center Frequency. This command will set the Center
Frequency to be used when the RF input is selected, even if the RF input is not the input that is
selected at the time the command is sent. Note that the Center Freq function in the Frequency menu
on the front panel always applies to the currently selected input.

Scope

Meas Global

Remote Command

[:SENSe]:FREQuency:RF:CENTer <freq>

[:SENSe]:FREQuency:RF:CENTer?

Example

FREQ:RF:CENT 30 MHz

Notes

This command is the same in all modes, but the parameter is Measurement Global. So the value

576

Remote Language Compatibility Measurement Application Reference

Advertising