Agilent Technologies Signal Analyzer N9030a User Manual

Page 165

Advertising
background image

3  Programming the Analyzer

STATus Subsystem

register bit has a positive transition (0 to 1). The variable <integer> is the sum of the decimal
values of the bits that you want to enable.

Mode

All

Remote
Command

:STATus:QUEStionable:INTegrity:SIGNal:PTRansition <integer>

:STATus:QUEStionable:INTegrity:SIGNal:PTRansition?

Example

STAT:QUES:INT:SIGN:PTR 4

Burst not found will be reported to the Integrity Summary of the Status Questionable register.

Preset

32767

Min

0

Max

32767

Status
Bits/OPC
dependencies

Sequential command

Initial S/W
Revision

Prior to A.02.00

Questionable Integrity Uncalibrated Register

The following commands and queries are available for this register:

"Questionable Integrity Uncalibrated Condition" on page 165

"Questionable Integrity Uncalibrated Enable" on page 166

"Questionable Integrity Uncalibrated Event Query" on page 166

"Questionable Integrity Uncalibrated Negative Transition" on page 167

"Questionable Integrity Uncalibrated Positive Transition" on page 167

Questionable Integrity Uncalibrated Condition

This query returns the decimal value of the sum of the bits in the Questionable Integrity
Uncalibrated Condition register.

The data in this register is continuously updated and reflects the current conditions.

Mode

All

Remote Command

:STATus:QUEStionable:INTegrity:UNCalibrated:CONDition?

Example

STAT:QUES:INT:UNC:COND?

Remote Language Compatibility Measurement Application Reference

165