6  RLC Swept SA Measurement Front-Panel & SCPI Reference
FREQ Channel

Preset and Max values are dependent on Hardware Options (5xx)

If no terminator (e.g. MHz) is sent the terminator Hz is used.  If a terminator with unit other than
Frequency is used, an invalid suffix error message is generated.

Dependencies

The Center Frequency can be limited by Start or Stop Freq limits, if the Span is so large that Start
or Stop reach their limit. 

Couplings

When operating in “swept span”, any value of the Center Frequency or Span that is within the
frequency range of the analyzer is allowed when the value is being set through the front panel
numeric key pad or the SCPI command. The other parameter is forced to a different value if
needed, to keep the Start and the Stop Frequencies within the analyzer’s frequency range

Preset

Depends on instrument maximum frequency, mode, measurement, and selected input.

See

"Center Frequency Presets" on page 574

and

"RF Center Freq" on page 576

and

Ext Mix

Center Freq

and 

"I/Q Center Freq" on page 578

.

State Saved

Saved in instrument state

Min

Depends on instrument maximum frequency, mode, measurement, and selected input..

See

"Center Frequency Presets" on page 574

and

"RF Center Freq" on page 576

and 

"I/Q Center

Freq" on page 578

.

Max

Depends on instrument maximum frequency, mode, measurement, and selected input..

See

"Center Frequency Presets" on page 574

and

"RF Center Freq" on page 576

and 

"I/Q Center

Freq" on page 578

.

Default Unit

Hz

Status Bits/OPC
Dependencies

Non-overlapped

Initial S/W Revision

Prior to A.02.00

Center Frequency Presets

The following table provides the Center Frequency Presets for the Spectrum Analyzer mode, and
the Max Freq, for the various frequency options: 

Freq Option

CF

after Mode Preset

Stop Freq after
Mode Preset

Max Freq

(can't
tune
above)

503
(all but N9000A)

1.805 GHz

3.6 GHz

3.7 GHz

574

Remote Language Compatibility Measurement Application Reference