Self test query, Wait-to-continue – Agilent Technologies Signal Analyzer N9030a User Manual

Page 110

Advertising
background image

3  Programming the Analyzer
IEEE 488.2 Common Commands

Remote Command

*TRG

Example

*TRG Triggers the instrument to take a sweep or start a measurement, depending on the current
instrument settings.

Notes

See related command :INITiate:IMMediate.

Initial S/W Revision

Prior to A.02.00

Self Test Query

This query performs the internal self-test routines and returns a number indicating the success of
the testing. A zero is returned if the test is successful, 1 if it fails.

Remote Command

*TST?

Example

*TST? Runs the self-test routines and returns 0=passed, 1=some part failed.

Initial S/W Revision

Prior to A.02.00

Wait-to-Continue

This command causes the instrument to wait until all overlapped commands are completed before
executing any additional commands. There is no query form for the command.

Remote Command

*WAI

Example

INIT:CONT OFF; INIT;*WAI Sets the instrument to single sweep. Starts a sweep and waits for its
completion.

Status Bits/OPC
dependencies

Not global to all remote ports or front panel. *OPC only considers operation that was initiated on
the same port as the *OPC command was issued from.

Initial S/W Revision

Prior to A.02.00

110

Remote Language Compatibility Measurement Application Reference

Advertising