Agilent Technologies Signal Analyzer N9030a User Manual

Page 983

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Recall

data. Pressing the Open key in this menu brings up the Open dialog and Open menu that
provides you with the options from where to recall the data. Once a filename has been selected or
entered in the Open menu, the recall occurs as soon as the Open button is pressed.

Key Path

Recall

Mode

All

Notes

The menu is built from whatever data types are available for the mode. Some keys will be
missing completely, so the key locations in the sub-menu will vary.

No SCPI command directly controls the Data Type that this key controls. The Data Type is
included in the MMEM:LOAD commands.

Dependencies

If a file type is not used by a certain measurement, it is grayed out for that measurement.  The
key for a file type will not show at all if there are no measurements in the Mode that support it.

Preset

Is not affected by Preset or shutdown, but is reset during Restore Mode Defaults

Readback

The data type that is currently selected

Initial S/W Revision

Prior to A.02.00

Amplitude Correction

This key selects the Amplitude Corrections as the data type to be imported.  When pressed a
second time, it brings up the Select Menu, which lets you select the Correction into which the data
will be imported.

Amplitude Corrections are fully discussed in the documentation of the Input/Output key, under
the Corrections key.

A set of preloaded Corrections files can be found in the directory

/My Documents/ EMC Limits and Ampcor. 

Under this directory, the directory called Ampcor (Legacy Naming) contains a set of legacy
corrections files, generally the same files that were supplied with older Agilent EMI analyzers, that
use the legacy suffixes .ant, .oth, .usr, and .cbl, and the old 8-character file names.  In the
directory called Ampcor, the same files can be found, with the same suffixes, but with longer,
more descriptive filenames.

When the Amplitude Correction is an Antenna correction and the Antenna Unit in the file is not
None

, the Y Axis Unit setting will change to match the Antenna Unit in the file.

Key Path

Recall

Mode

SA|EDGEGSM|PN

Remote Command

:MMEMory:LOAD:CORRection 1|2|3|4|5|6, <filename>

Remote Language Compatibility Measurement Application Reference

983

Advertising