Agilent Technologies Signal Analyzer N9030a User Manual

Page 1220

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping
or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger
event occurs and all the applicable trigger criteria have been met). A corresponding pop-up
message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2
sec. This message goes away when a trigger signal appears.

Initial S/W Revision

Prior to A.02.00

Trigger Level

Sets the value where the external 1 trigger input will trigger a new sweep/measurement.

Key Path

Trigger, External 1

Remote Command

:TRIGger[:SEQuence]:EXTernal1:LEVel <level>

:TRIGger[:SEQuence]:EXTernal1:LEVel?

Example

TRIG:EXT1:LEV 0.4 V

Couplings

This same level is used for the Ext1 trigger source in the Trigger menu, for the Ext1 selection in
the Periodic Timer sync source (in the Trigger menu and in the Gate Source menu), and also for
the Ext1 selection in the Gate Source menu.

Preset

1.2 V

State Saved

Saved in instrument state

Min

–5 V

Max

5 V

Default Unit

V

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:EXTernal:LEVel

For backward compatibility, the parameter EXTernal is mapped to EXTernal1

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:FRAMe:EXTernal1:LEVel

Initial S/W Revision

Prior to A.02.00

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on
a falling edge.

Key Path

Trigger, External 1

Remote Command

:TRIGger[:SEQuence]:EXTernal1:SLOPe POSitive|NEGative

1220

Remote Language Compatibility Measurement Application Reference

Advertising