6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output

Example

SENS:CORR:CSET1 ON

Dependencies

Turning this on automatically turns on "Apply Corrections"

Only the first correction array (Correction 1) supports antenna units.  When this array is turned
on, and it contains an Antenna Unit other than “None”, the Y Axis Unit of the analyzer is forced
to that Antenna Unit. All other Y Axis Unit choices are grayed out.  

Note that this means that a correction file with an Antenna Unit can only be loaded into the
Corrections 1 register.  Consequently only for Correction 1 does the dropdown in the Recall dialog
include.ant, and if an attempt is made to load a correction file into any other Correction register
which DOES contain an antenna unit, a Mass Storage error is generated. 

This command will generate an “Option not available” error unless you have the proper option
installed in your instrument. 

Preset

Not affected by a Preset. Set to OFF by Restore Input/Output Defaults

State Saved

Saved in instrument state.

Backwards Compatibility
Notes

Unlike legacy analyzers, Preset does not turn Corrections off (Restore Input/Output Defaults
does).

Initial S/W Revision

A.02.00

Properties

Accesses a menu that lets you set the properties of the selected correction.

Key Path

Input/Output, Corrections

Initial S/W Revision

A.02.00

Select Correction

Specifies the selected correction. The term "selected correction" is used throughout this document
to specify which correction will be affected by the functions.

Key Path

Input/Output, Corrections

Mode

SA

Notes

The selected correction is remembered even when not in the correction menu.

Preset

Set to Correction 1 by Restore Input/Output Defaults

Readback

Correction 1|Correction 2|Correction 3|Correction 4|Correction 5|Correction 6

Initial S/W Revision

A.02.00

628