Fa (start frequency), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual

Page 291: Description, Syntax legacy products

Advertising
background image

5  Legacy Command Descriptions

FA (Start Frequency)

FA (Start Frequency)

Syntax

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

Specifies the start frequency value. The start frequency is equal to the center frequency minus the
span divided by two (FA = CF – SP/2). Changing the start frequency changes the center frequency
and span.

Format

FA <real>HZ|KHZ|MHZ|GHZ|KZ|MZ|GZ

<real>: Default unit is Hz. Range: frequency range of the instrument.

FA UP|DN

Step Increment: Frequency span divided by 10.

FA OA

Specifying OA returns only the current value to the controller. It does not set the active

Remote Language Compatibility Measurement Application Reference