5 Legacy Command Descriptions
FA (Start Frequency)
FA (Start Frequency)
Syntax
Legacy Products
8560 series, 8566A/B, 8568A/B
Description
Specifies the start frequency value. The start frequency is equal to the center frequency minus the
span divided by two (FA = CF – SP/2). Changing the start frequency changes the center frequency
and span.
Format
FA <real>HZ|KHZ|MHZ|GHZ|KZ|MZ|GZ
<real>: Default unit is Hz. Range: frequency range of the instrument.
FA UP|DN
Step Increment: Frequency span divided by 10.
FA OA
Specifying OA returns only the current value to the controller. It does not set the active
Remote Language Compatibility Measurement Application Reference