6  RLC Swept SA Measurement Front-Panel & SCPI Reference
System

If the value of the Timebase DAC changes (by switching to Calibrated from User with User set to a
different value, or in User with a new value entered) an alignment may be necessary. The
alignment system will take appropriate action; which will either invoke an alignment or cause an
Alert.

Key Path

System, Alignments

Mode

All

Remote Command

:CALibration:FREQuency:REFerence:MODE CALibrated|USER

:CALibration:FREQuency:REFerence:MODE?

Example

:CAL:FREQ:REF:MODE CAL

Notes

If the value of the timebase is changed the alignment system automatically performs an
alignment or alerts that an alignment is due.

If the value of the timebase is changed the alignment system automatically performs an
alignment or alerts that an alignment is due.

Preset

This is unaffected by Preset but is set to CALibrated on a “Restore System Defaults->Align”.

State Saved

No

Initial S/W Revision

Prior to A.02.00

Calibrated

Sets the Timebase DAC to the value established during factory or field calibration. The value
displayed on the menu key is the calibrated value.

Key Path

System, Alignments, Timebase DAC

Mode

All

Example

:CAL:FREQ:REF:MODE CAL

Readback Text

[xxx]

< where xxx is the calibrated value

Initial S/W Revision

Prior to A.02.00

User

Allows setting the Timebase DAC to a value other than the value established during the factory or
field calibration. The value displayed on the menu key is the calibrated value.

Key Path

System, Alignments, Timebase DAC

1292

Remote Language Compatibility Measurement Application Reference