Channel (channel selection), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual

Page 259: Description, Syntax legacy products

Advertising
background image

5  Legacy Command Descriptions

CHANNEL (Channel Selection)

CHANNEL (Channel Selection)

Syntax

Legacy Products

8560 series

Description

Increments or decrements the instrument center frequency by one channel spacing.

The channel spacing value is set using the command

"ACPSP (Adjacent Channel Power Channel

Spacing)" on page 228

.

Format

CHANNEL UP|DN

Query Data Type

N/A

SCPI Equivalent Commands

[:SENSe]:FREQuency:CENTer <freq>

[:SENSe]:FREQuency:CENTer?

(See

"Center Freq" on page 573

)

Remote Language Compatibility Measurement Application Reference