Agilent Technologies Signal Analyzer N9030a User Manual

Page 604

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output

be identical. This does not apply to Probe settings or to parameters that are determined by the
probe.

Key Path

Input/Output, I/Q, Q Setup

Remote Command

:INPut:IQ:MIRRored OFF|ON|0|1

:INPut:IQ:MIRRored?

Example

Turn off the mirroring of parameters from I to Q.

INP:IQ:MIRR OFF

Couplings

Only displayed for the Q channel. When Yes, the I channel values for some parameters are
mirrored (copied) to the Q channel. However, when a parameter is determined by the type of
probe and a probe is sensed, the probe setting is always used and the I channel setting is
ignored. The following parameters are mirrored:

Differential Input (when not determined by probe)

Input Z (when not determined by probe)

Preset

This is unaffected by a Preset but is set to the default value (Q Same as I set to "On") on a
"Restore Input/Output Defaults" or "Restore System Defaults->All"

State Saved

Saved in instrument state.

Range

On | Off

Readback Text

"Q Same as I" when On, otherwise none.

Initial S/W Revision

Prior to A.02.00

Q Differential Input

Selects differential input on or off for the Q channel. For differential input (also called balanced
input), the analyzer uses both the Q and Q-bar ports. When differential input is off (also called
single-ended or unbalanced input), the analyzer uses only the Q port.

Key Path

Input/Output, I/Q, Q Setup

Remote Command

:INPut:IQ:Q:DIFFerential OFF|ON|0|1

:INPut:IQ:Q:DIFFerential?

Example

Put the Q channel in Differential Input mode

INP:IQ:Q:DIFF ON

Notes

When Differential Input = On, the analyzer will check for attenuation mismatches between the Q
and Q-bar ports. If the difference in attenuation values exceeds 0.5 dB a Settings Alert error
condition, error  159 will be set.

When Q Differential Input = On, and IQ Path is I+jQ, the I Differential input must also be On. 
Similarly, when Q Differential Input = Off, and IQ Path is I+jQ, the I Differential input must also
be Off.  If the states of the two inputs do not match, an error condition message is generated,

604

Remote Language Compatibility Measurement Application Reference

Advertising