3  Programming the Analyzer

STATus Subsystem

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Integrity Uncalibrated Negative Transition

This command determines which bits in the Questionable Integrity Uncalibrated Condition
register will set the corresponding bit in the Questionable Integrity Uncalibrated Event register
when the condition register bit has a negative transition (1 to 0). The variable <integer> is the
sum of the decimal values of the bits that you want to enable.

Mode

All

Remote
Command

:STATus:QUEStionable:INTegrity:UNCalibrated:NTRansition <integer>

:STATus:QUEStionable:INTegrity:UNCalibrated:NTRansition?

Example

STAT:QUES:INT:UNC:NTR 1

Oversweep cleared will be reported to the Integrity Summary of the Status Questionable register.

Preset

0

Min

0

Max

32767

Status
Bits/OPC
dependencies

Sequential command

Initial S/W
Revision

Prior to A.02.00

Questionable Integrity Uncalibrated Positive Transition

This command determines which bits in the Questionable Integrity Uncalibrated Condition
register will set the corresponding bit in the Questionable Integrity Uncalibrated Event register
when the condition register bit has a positive transition (0 to 1). The variable <integer> is the
sum of the decimal values of the bits that you want to enable.

Mode

All

Remote
Command

:STATus:QUEStionable:INTegrity:UNCalibrated:PTRansition <integer>

:STATus:QUEStionable:INTegrity:UNCalibrated:PTRansition?

Example

STAT:QUES:INT:UNC:PTR 1

Oversweep (Meas Uncal) occurred will be reported to the Integrity Summary of the Status Questionable

Remote Language Compatibility Measurement Application Reference