Agilent Technologies Signal Analyzer N9030a User Manual

Page 1419

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trigger

to OFF, and the external sync source rate is changed for some reason, the periodic timer is
synchronized at the every external synchronization pulse by resetting the internal state of the
timer circuit.

Key Path

Trigger, Periodic Timer

Remote Command

:TRIGger[:SEQuence]:FRAMe:PERiod <time>

:TRIGger[:SEQuence]:FRAMe:PERiod?

Example

TRIG:FRAM:PER 100 ms

Dependencies

The invalid data indicator turns on when the period is changed, until the next
sweep/measurement completes.

Couplings

The same period is used in the Gate Source selection of the period timer.

Preset

20 ms

GSM: 4.615383

State Saved

Saved in instrument state

Min

100.000 ns

Max

559.0000 ms

Default Unit

S

Initial S/W Revision

Prior to A.02.00

Offset

Adjusts the accumulated offset between the periodic timer events and the trigger event. Adjusting
the accumulated offset is different than setting an offset, and requires explanation.

The periodic timer is usually not synchronized with any external events, so the timing of its
output events has no absolute meaning. Since the timing relative to external events (RF signals) is
important, you need to be able to adjust (offset) it. However, you have no direct way to see when
the periodic timer events occur. All that you can see is the trigger timing. When you want to
adjust the trigger timing, you will be changing the internal offset between the periodic timer
events and the trigger event. Because the absolute value of that internal offset is unknown, we will
just call that the accumulated offset. Whenever the Offset parameter is changed, you are changing
that accumulated offset. You can reset the displayed offset using Reset Offset Display. Changing
the display does not change the value of the accumulated offset, and you can still make
additional changes to accumulated offset.

To avoid ambiguity, we define that an increase in the "offset" parameter, either from the knob or
the SCPI adjust command, serves to delay the timing of the trigger event.

Key Path

Trigger, Periodic Timer