5 Legacy Command Descriptions
KSZ (Reference Level Offset)
KSZ (Reference Level Offset)
Syntax
Legacy Products
8566A/B, 8568A/B
Description
Offsets all amplitude readouts on the display but without affecting the trace.
Once activated, KSZ displays the amplitude offset on the left side of the screen.
Sending KSZ 0, or presetting the instrument, eliminates an amplitude offset.
Format
KSZ <real>DM|MV|UV
8566A/B only supports unit DM
KSZ OA
KSZ?
Query Data Type
<real>
SCPI Equivalent Commands
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet <rel_ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet?
(See
Preset
0
Notes
The functions of KSZ are identical to
"ROFFSET (Reference Level Offset)" on page 452